Transformer detection High voltage withstand test equipment Switch detection equipment electric energy measurement and testing equipment secondary circuit test equipment Lightning arrester/insulator testing equipment Ground/Insulation Resistance Test Equipment SF6Gas detection category Phase voltammeter Frequency conversion grounding grid grounding resistance tester Digital intelligent partial discharge tester Vacuum degree Micro water meter Insulating oil dielectric strength tester Insulating oil micro-water tester Wired vector Flatbed Vector Analyzer Taiqu District User Identifier Cable fault detector Generator rotor AC impedance tester Dielectric loss tester Nuclear phase analyzer Electric energy meter field calibrator Underground pipeline detector AJRZTransformer short circuit impedance tester Oil and water tester Transformer short circuit impedance tester DC resistance tester Generator AC impedance Battery testing equipment
AJZC Direct resistance tester
Functional Features
1. The whole machine is controlled by a high-speed microcontroller, with a high degree of automation and easy operation.
2. The instrument adopts a new power technology and has a wide measurement range.
3. The protection function is complete, which can reliably protect the impact of back electromotive force on the instrument, and the performance is more reliable.
4. Fast response speed, stable instrument measurement data, and automatic data refreshing during instrument testing.
6. Intelligent power management technology ensures that instruments always operate at their minimum power, effectively reducing internal heat generation and saving energy.
7. The instrument is equipped with a non power-off clock inside.
8. The instrument has an internal non power down memory that can permanently store data.
Technical indicators
1. Output current: automatic, 10A, 5A, 1A, 200mA, 40mA,<5mA
2. Resolution: 0.1 μ Ω
3. Range: 100 Ω -100K Ω (<5mA range)
1 Ω -500 Ω (40mA range)
100m Ω -100 Ω (200mA range)
5m Ω -20 Ω (1A gear)
1m Ω -4 Ω (5A gear)
0.5m Ω -2 Ω (10A gear)
4. Accuracy: 0.2%
5. Working temperature: 0-40 ℃
6. Working humidity:<90% RH, no condensation